Hours - Contract Technician4 69.00 276.004/8/2024
Clean, calibrate and certify West Influent
flowmeter and West Supernate flowmeter.
Xmtr Fee - Contract Transmitter Certification Fee2 49.00 98.00
Fuel Surcharge - ... Broward and other1 10.00 10.00
$384.00
$384.00
$0.00
P .C. Controls Inc PC
CONTROLS
4114 W Palm Aire Dr 14 IB , Pompano Beach, FL 33069
Phone : (954) 568-9663, Fax: TBD
J: r--..:r c::: E-mail: pccontrolsinc @ bellsouth.net
Certified Flow Meter Test Record
Customer: City of Dania Beach
Consumptive Use Permit # 06-00187-W
Test Site West Influent Flow Transmitter District ID#
Contact Nate Costa Number 954 924-3616
Meter Model Seametrics FT440W Serial # 06201036
Meter Accessories: None
Pipe Information at test site
Material Ductile Iron
Outside Diameter 13 .2" OD Pipe Thickness .4" WT
Transducer Spacing 9.99" V-Mount Distance from Meter installation is
Test Information using Fuji Porta-Flow Portable Ultrasonic Flowmeter
Test perfonned @ __ 1250.0 ____ GPM
Initial Meter Reading at Calibration Start: 740000 X 1 End : 752000 x 1
_12 ,000 __ Gallons Totalized in _574 __ Seconds Average _1254.4_ GPM
Test Meter Reading at Calibration Start: 0 X I End : 12000 __ X _I _
12,000 __ Gallons Totalized in _ 600 __ Seconds Average _1250 .0_ GPM
Meter Tested at _99.6_ % Accuracy Last Meter Test N I A Last Accuracy __ _
Comments: Prior to calibration meter was reading 99.6% oftest meter -Site meter read 1254.4 GPM and
test meter read 1250.0 GPM.
Performed By: NP Conaway Date: 4/8/24
Thi s is to certify that the above li sted equipment has been c ibrated and mee ts or exceeds all publi shed s pecification s.
Calibration was performed using standard s and/or in st entation who se accuracy is traceabl e to the N .l.S.T.
Signature
PC
CONTROLS
J: r--..:r c::::
P.C. Controls Inc
4114 W Palm Aire Dr 14 IB, Pompano Beach, FL 33069
Phone: (954) 568-9663, Fax: TBD
E-mail: pccontrolsinc@bellsouth.net
Certified Flow Meter Test Record
Customer: City of Dania Beach
Consumptive Use Permit# 06-00187-W
Test Site West Supematent Flow Transmitter, FIT-103 District ID# ___ _
Contact Nate Costa Number 954 924-3616
Meter Model Seametrics FT440W Serial# 042017000073
Meter Accessories: None
Pipe Information at test site
Material PVC Schedule 40
Outside Diameter 2.375" OD Pipe Thickness .154" WT
Transducer Spacing 0 .55" V-Mount Distance from Meter installation is 3 feet
Test Information using Fuji Porta-Flow Portable Ultrasonic Flowmeter
Test performed @ __ 26.2 _____ GPM
lnitial Meter Reading at Calibration Start: 2005 X 1 End: 2305 X I ------
300 Gallons Totalized in 679 Seconds ---Average _ 26 .5_ GPM
Test Meter Reading at Calibration Start: 0 Xl ---End: 300 X I ---
300 Gallons Totalized in 688 Seconds ---Average _26.2_ GPM
Meter Tested at _I 00.l_ % Accuracy Last Meter Test N I A ---Last Accuracy __ _
Comments: Prior to calibration meter was reading 10 I . I% of test meter -Site meter read 26.5 GPM and
test meter read 26.2 GPM.
Performed By: NP Conaway Date: 4/8/24
This is to certify that the above listed equipment has been calibrated and meets or exceeds all published specifications.
Calibration was performed using standards and/or instrumentation w ose accuracy is traceable to the N.l.S.T.
Signature